Power And DC Bias Enhance
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Entry model
No Interface
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1USB Software I/O Libraries
Qty: 1Power and DC Bias enhance (20V, 100mA, 40V DC Bias, 2nd DC source)
Bias Current Interface
No Interface
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Power and DC Bias enhance (20V, 100mA, 40V DC Bias, 2nd DC source)
No Interface
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Power and DC Bias enhance (20V, 100mA, 40V DC Bias, 2nd DC source)
No Interface
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Power and DC Bias enhance (20V, 100mA, 40V DC Bias, 2nd DC source)
Scanner Interface
No Interface
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1DCR Measurement
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Power and DC Bias enhance (20V, 100mA, 40V DC Bias, 2nd DC source)
Handler interface
Scanner Interface
Cord Power
Qty: 1Cable USB-A - USB-MiniB
Qty: 1Handle
Qty: 1USB Drive with Software and Manuals
Qty: 1Keysight Technologies ESG User Guide
Content is AI-generated. Verify critical details with manufacturer-provided resources.
The Keysight E4980A precision LCR meter covers a frequency range from 20 Hz to 2 MHz with 0.01 Hz resolution at low frequencies. The test frequencies can be set in fine increments: 0.01 Hz steps from 20 to 99.99 Hz, 0.1 Hz from 100 to 999.9 Hz, and up to 1 kHz steps at the highest frequencies.
The E4980A delivers exceptional measurement accuracy with basic accuracy as low as 0.05% in medium and long measurement time modes. For example, at 1 kHz with 1 Vrms test signal, measuring a 100 nF capacitor achieves 0.05% capacitance accuracy. The absolute accuracy combines basic accuracy, calibration accuracy, and relative accuracy components including temperature coefficient and offset corrections.
The E4980A measures a comprehensive range of parameters including Cp-D, Cp-Q, Cp-G, Cp-Rp, Cs-D, Cs-Q, Cs-Rs, Lp-D, Lp-Q, Lp-G, Lp-Rp, Lp-Rdc, Ls-D, Ls-Q, Ls-Rs, Ls-Rdc, R-X, Z with phase angle (degree/radian), G-B, Y with phase angle, and Vdc-Idc. It supports both parallel and series equivalent circuit models for flexible component characterization.
The E4980A provides test signal voltage from 0 to 2.0 Vrms with resolution from 100 uVrms to 1 mVrms depending on the range, and test signal current from 0 to 20 mArms with resolution from 1 to 10 uArms. Both normal and constant signal modes are available, with constant mode maintaining selected voltage or current at the device under test independently of impedance changes.
Yes, the E4980A provides DC bias voltage from 0 V to +2 V in three discrete steps (0 V, 1.5 V, 2 V) with accuracy of 0.1% + 2 mV at 23 degrees C. The DC bias enhancement option (E4980A-001) enables extended capabilities for testing components under realistic operating conditions, which is particularly important for capacitors and magnetic components.
The E4980A list sweep function supports up to 201 measurement points. The primary sweep parameter can be test frequency, test signal voltage, test signal current, DC bias voltage, DC bias current, or DC source voltage. A secondary parameter can also be configured via SCPI commands, enabling comprehensive multi-dimensional component characterization.
The E4980A offers three types of compensation: OPEN compensation for stray admittance (C, G) of test fixtures, SHORT compensation for residual impedance (L, R) of fixtures, and LOAD compensation for errors between actual measured values and known standard values. These compensations ensure accurate measurements by removing parasitic effects from cables and fixtures.
The E4980A features a data buffer function that stores up to 201 measurement results for batch readout. The save/recall function allows storage of up to 10 setup conditions in built-in non-volatile memory and an additional 10 setups on USB memory. It supports USB 1.1 mass storage class devices with FAT16/FAT32 format and includes auto-recall capability.