Time Domain Analysis
2-port Test Set, 100 Khz To 18 Ghz
Cord Power
Qty: 1Add Handler IO
2-port test set, 100 kHz to 18 GHz
Add GPIB
Cord Power
Qty: 1Time Domain Analysis
2-port Test Set, 100 Khz To 18 Ghz
Add Gpib
Add Handler Io
Cord Power
Qty: 12-port Test Set, 100 Khz To 18 Ghz
Time Domain Analysis
Not Required Gpib
Not Required Handler Io
Cord Power
Qty: 1Cord Power
Qty: 1Time domain analysis/Test Wizard
2-port test set, 100 kHz to 18 GHz
Add GPIB
Add Handler IO
Cord Power
Qty: 1Wireless power transfer analysis
Time domain analysis/Test Wizard
2-port test set, 100 kHz to 18 GHz
Add GPIB
Add Handler IO
Cord Power
Qty: 1Content is AI-generated. Verify critical details with manufacturer-provided resources.
The E5063A-2H5 covers 100 kHz to 18 GHz, making it well suited for characterizing components and devices across commercial wireless, satellite, radar, and microwave frequency bands. The analyzer can actually be set from 50 kHz with typical performance data available at that range, though warranted specifications apply from 100 kHz onward.
With 3 kHz IF bandwidth, the E5063A-2H5 provides 92 dB dynamic range from 100 MHz to 4.34 GHz, 81 dB from 4.34 to 8.5 GHz, 75 dB from 8.5 to 13 GHz, 65 dB from 13 to 16 GHz, and 62 dB from 16 to 18 GHz. Switching to 10 Hz IF bandwidth improves these figures to 117 dB, 106 dB, 100 dB, 90 dB, and 87 dB respectively, enabling precise characterization of low-loss and high-isolation devices.
The E5063A-2H5 supports the 85032F (Type-N, 50 ohm), 85033E (3.5 mm), and 85052D (3.5 mm) mechanical calibration kits, plus the 85092C, 85093C, and N4691B electronic calibration modules. Using the 85052D with full 2-port cal, directivity reaches 42 dB below 10 MHz and 36 dB from 10 to 18 GHz. Electronic cal modules like the N4691B achieve up to 48 dB directivity in the mid-band.
The E5063A-2H5 has a nominal preset output of -5 dBm. The adjustable range is -20 to 0 dBm from 100 kHz to 300 kHz, -20 to -5 dBm from 300 kHz to 8.5 GHz, and -15 to -5 dBm from 8.5 to 18 GHz, all with 0.05 dB resolution. Level accuracy at 50 MHz is ±0.9 dB absolute, while flatness across 1 MHz to 4.34 GHz is within ±1.0 dB relative to 50 MHz.
Transmission trace noise at maximum output power is 2 mdB rms from 8.5 MHz to 4.34 GHz (70 kHz IFBW), rising to 20 mdB rms from 16 to 18 GHz. Phase trace noise is 0.015 deg rms in the best band and 0.13 deg rms from 16 to 18 GHz. Magnitude stability is ±0.01 dB/°C from 300 kHz to 6 GHz and ±0.04 dB/°C from 12 to 18 GHz, ensuring reliable measurements across changing lab temperatures.
The E5063A-2H5 test port maximum input level is +6 dBm during normal operation. The damage level threshold is +26 dBm or ±35 VDC, providing substantial margin for accidental overloads. The receiver compression at maximum input level is ±0.2 dB in magnitude across most of the frequency range, with phase compression of ±1.5 deg from 1 MHz to 4.34 GHz.
Measurement throughput depends on span, point count, calibration type, and IF bandwidth. For a 100 kHz to 18 GHz sweep with 201 points, a 2-port calibrated S21 measurement completes in about 50 ms at 300 kHz IFBW, 74 ms at 30 kHz IFBW, and 820 ms at 1 kHz IFBW. Data transfer over 100 Mbps LAN socket takes just 1 ms for 201 points in REAL 64 format.
The E5063A-2H5 rear panel includes GPIB (IEEE-488), USB host ports, USBTMC interface (USB 2.0), 10/100/1000 BaseT Ethernet LAN, VGA video output, a 36-pin handler I/O port, and BNC connectors for external trigger input/output and 10 MHz reference input/output. The reference output provides a 10 MHz sinewave at 0 dBm ±3 dB into 50 ohm. Line power accepts 90 to 264 VAC at 47 to 63 Hz.